Autor: |
D. G. Sarantites, N. Mullani, L. Westerberg, J. H. Barker, R. Lovett, C. M. Currie, J.T. Hood |
Rok vydání: |
1977 |
Předmět: |
|
Zdroj: |
Nuclear Instruments and Methods. 145:295-310 |
ISSN: |
0029-554X |
DOI: |
10.1016/0029-554x(77)90425-6 |
Popis: |
Expressions for interpreting γ-ray multiplicity measurements with a Ge(Li) and N NaI detectors are derived. A scattering chamber designed to allow measurements of multiple coincidences between a Ge(Li) detector and/or particle detectors and up to 12 NaI detectors is described. A special multiplicity recording unit has been constructed. It employs the strobed overlap coincidence technique and it is suitable for routing a multichannel analyzer or for interfacing to a computer. Details of the analysis procedure are discussed. The application of a computer code to multiplicity calculations is described. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|