A statistical approach to infer the minimum setup distance of a well chamber to the wall or to the floor for Ir192 HDR calibration

Autor: Chen Shou Chui, Jeng Hung Lee, Liyun Chang, Sheng Yow Ho, Yi Chun Du, Tainsong Chen
Rok vydání: 2008
Předmět:
Zdroj: Medical Physics. 35:2214-2217
ISSN: 0094-2405
DOI: 10.1118/1.2911156
Popis: We propose a new method based on statistical analysis technique to determine the minimum setup distance of a well chamber used in the calibration of Ir 192 high dose rate (HDR). The chamber should be placed at least this distance away from any wall or from the floor in order to mitigate the effect of scatter. Three different chambers were included in this study, namely, Sun Nuclear Corporation, Nucletron, and Standard Imaging. The results from this study indicated that the minimum setup distance varies depending on the particular chamber and the room architecture in which the chamber was used. Our result differs from that of a previous study by Podgorsak et al. [Med. Phys.19, 1311–1314 (1992)], in which 25 cm was suggested, and also differs from that of the International Atomic Energy Agency (IAEA)-TECDOC-1079 report, which suggested 30 cm . The new method proposed in this study may be considered as an alternative approach to determine the minimum setup distance of a well-type chamber used in the calibration of Ir 192 HDR.
Databáze: OpenAIRE