Measurement uncertainty of a measurement system for dynamic pressure in the kbar regime
Autor: | O Slanina, R Wynands |
---|---|
Rok vydání: | 2021 |
Předmět: | |
Zdroj: | Measurement Science and Technology. 32:075010 |
ISSN: | 1361-6501 0957-0233 |
DOI: | 10.1088/1361-6501/abe47e |
Popis: | We have characterized the measurement uncertainty of a setup for the dynamic measurement of pressure pulses with amplitudes in the low kbar range and with millisecond duration. The uncertainty is closely proportional to pressure, with a magnitude of 0.34% (coverage factor k = 1). In particular, we treat the safety-critical application of measuring the pressure inside an ammunition cartridge during firing, where a target uncertainty of 3% has been set by the Permanent International Commission for the Proof of Small Arms. |
Databáze: | OpenAIRE |
Externí odkaz: |