Multiple inadäquate Defibrillatorschocks bei Isolationsdefekt einer Biotronik-Linox®-Defibrillatorelektrode mit externalisiertem Leiter
Autor: | Wolfram Grimm, Hamdi Elfarra, L. Rybinski, Rainer Moosdorf |
---|---|
Rok vydání: | 2016 |
Předmět: |
medicine.medical_specialty
business.industry INSULATION FAILURE 030204 cardiovascular system & hematology medicine.disease Treatment failure Sudden cardiac death 03 medical and health sciences Equipment failure 0302 clinical medicine Device removal Physiology (medical) Internal medicine medicine Cardiology 030212 general & internal medicine Cardiology and Cardiovascular Medicine business Defibrillator lead Lead (electronics) |
Zdroj: | Herzschrittmachertherapie + Elektrophysiologie. 27:70-72 |
ISSN: | 1435-1544 0938-7412 |
DOI: | 10.1007/s00399-016-0418-6 |
Popis: | In this article the case of a patient who received a total of 35 inappropriate defibrillator shocks due to insulation failure with externalized conductor of a Biotronik Linox® lead is described. The implanted defibrillator was immediately inactivated and the failed lead was extracted using a laser sheath system. |