Development of a mobile radon calibration system at KRISS

Autor: Byungjoo Kim, M.J. Han, Dong-Hye Heo, B. C. Kim, J.M. Lee, K.B. Lee, Y. G. Seon, Seung-Mi Hwang
Rok vydání: 2021
Předmět:
Zdroj: Journal of Radioanalytical and Nuclear Chemistry. 330:571-576
ISSN: 1588-2780
0236-5731
DOI: 10.1007/s10967-021-07932-4
Popis: A mobile radon calibration system was developed with a high-sensitivity radon detector, a movable calibration chamber, and a 226Ra solid source at the Korea Research Institute of Standards and Science (KRISS). The high-sensitivity radon detector was based on electrostatic collection mechanism. The volume of the calibration chamber was (1003.8 ± 6.0) L. The mobile radon calibration system was calibrated with 222Rn CRM from the radon primary system at KRISS. The sensitivities were established to be (231.8 ± 16.1) cpm/(kBq/m3) for 218Po and (254.9 ± 17.8) cpm/(kBq/m3) for 214Po, respectively (k = 2).