Development of a mobile radon calibration system at KRISS
Autor: | Byungjoo Kim, M.J. Han, Dong-Hye Heo, B. C. Kim, J.M. Lee, K.B. Lee, Y. G. Seon, Seung-Mi Hwang |
---|---|
Rok vydání: | 2021 |
Předmět: |
Health
Toxicology and Mutagenesis Detector Public Health Environmental and Occupational Health chemistry.chemical_element Radon Pollution Analytical Chemistry Nuclear Energy and Engineering chemistry Calibration Environmental science Radiology Nuclear Medicine and imaging Spectroscopy Remote sensing |
Zdroj: | Journal of Radioanalytical and Nuclear Chemistry. 330:571-576 |
ISSN: | 1588-2780 0236-5731 |
DOI: | 10.1007/s10967-021-07932-4 |
Popis: | A mobile radon calibration system was developed with a high-sensitivity radon detector, a movable calibration chamber, and a 226Ra solid source at the Korea Research Institute of Standards and Science (KRISS). The high-sensitivity radon detector was based on electrostatic collection mechanism. The volume of the calibration chamber was (1003.8 ± 6.0) L. The mobile radon calibration system was calibrated with 222Rn CRM from the radon primary system at KRISS. The sensitivities were established to be (231.8 ± 16.1) cpm/(kBq/m3) for 218Po and (254.9 ± 17.8) cpm/(kBq/m3) for 214Po, respectively (k = 2). |