Development, calibration and application of new generation dissector with picosecond temporal resolution

Autor: D. V. Shvedov, A V Smirnov, Victor Dorokhov, E.I. Zinin, M. Ya. Schelev, A. K. Vereshchagin, E. V. Shashkov, O. V. Anchugov, Oleg Meshkov, A. N. Petrozhitsky, A. I. Zarovskii, G.Ya. Kurkin, P. B. Gornostaev
Rok vydání: 2017
Předmět:
Zdroj: SPIE Proceedings.
ISSN: 0277-786X
Popis: A dissector is an electron-optical device designed for measurement of periodic light pulses of subnanosecond and picosecond duration. LI-602 dissector developed at Budker Institute of Nuclear Physics (BINP SB RAS) is widely used for routine measurements of a longitudinal profile of electron and positron beams at BINP electron-positron colliders and other similar installations (1,2) . LI-602 dissector is a part of many optical diagnostic systems and provides temporal resolution of about 20 ps. Recently a new generation of picosecond dissectors were created on the basis of the PIF- 01/S1 picosecond streak-image tube designed and manufactured at the A.M.Prokhorov General Physics Institute (GPI) Photoelectronics Department (3,4) . The results of the measurements of instrument function of the new dissector based on PIF-01/S1, which were carried out in the static mode (5) showed that temporal resolution of the dissector can be better than 3-4 ps (FWHM). The results of temporal resolution calibration of the new-generation picosecond dissector carried out at the specialized set-up based on a femtosecond Ti:sapphire laser and recent results of longitudinal beam profile measurements at BINP damping ring are given in this work.
Databáze: OpenAIRE