Calculation of the absolute peak efficiency of Ge and Ge(Li) detectors for different counting geometries

Autor: A. Simonits, J. De Donder, F. De Corte, A. De Wispelaere, Lin Xilei, Luc Moens, J. Hoste
Rok vydání: 1982
Předmět:
Zdroj: Journal of Radioanalytical Chemistry. 70:539-550
ISSN: 1588-2780
0134-0719
DOI: 10.1007/bf02516136
Popis: A new method is presented to calculate with improved accuracy the absolute peak efficiency of cylindrical Ge and Ge(Li) detectors for point, disk and cylinder sources, positioned at any source-detector distance. Moreover attention was paid to true-coincidence effects. The method is extensively tested and applied for the analysis of reference materials. The accuracy turned out to be 3% or better.