Difference measurement using Coriolis mass flowmeters

Autor: W. Kiehl
Rok vydání: 1991
Předmět:
Zdroj: Flow Measurement and Instrumentation. 2:135-138
ISSN: 0955-5986
DOI: 10.1016/0955-5986(91)90023-k
Popis: When Coriolis mass flow meters are used close to each other in applications such as leak detection, difference measurements or parallel flow configurations, the specific features of the transducers can lead to reduced accuracy and repeatability. The results of employing special evaluation procedures in combination with tuning the dynamic properties of the transducers to reduce such problems are illustrated by an example in which accuracy, repeatability and a high degree of synchronism of two adjacent meters are the main requirements.
Databáze: OpenAIRE