Achievement of the analytically ideal steady-state response at a microelectrode-based scanning electrochemical detector under flow injection analysis and normal-phase chromatography conditions
Autor: | Barry L. Reed, Russell John Tait, Barrie C. Finnin, Alan M. Bond, Peter C. Bury |
---|---|
Rok vydání: | 1993 |
Předmět: | |
Zdroj: | Analytical Chemistry. 65:3252-3257 |
ISSN: | 1520-6882 0003-2700 |
DOI: | 10.1021/ac00070a014 |
Popis: | A study of the one-electron reversible ferrocene oxidation process has been made under both flow injection analysis and normal-phase chromatography conditions in a thin-layer flow cell with a rapid-scanning microelectrode-based detector. A flow rate of 1.0 mL min -1 , scan rates in the range of 2-10 V s -1 , and platinum disk microelectrodes of radii in the range 5-50 μm enable voltammograms with the theoretically predicted steady-state caracteristics to be obtained. Ready access to the analytically ideal steady-state response and to electrode process characterization coupled with other advantages available with microelectrodes provides the rapid-scanning microelectrode-based detector with substantial advantages over many of the more commonly used forms of electrochemical detection |