Theoretical Research and Development of a Clinical Setup Prototype for Online Monitoring of the Bragg Peak Position for the Prometheus Proton Therapy Complex
Autor: | Yu. V. Altukhov, I. I. Degtyarev, A. A. Pryanichnikov, F. N. Novoskoltsev, E. V. Altukhova, V. E. Balakin |
---|---|
Rok vydání: | 2018 |
Předmět: |
Physics
Nuclear and High Energy Physics Radiation Proton 010308 nuclear & particles physics business.industry Physics::Medical Physics Monte Carlo method Gamma ray Bragg peak 01 natural sciences Atomic and Molecular Physics and Optics Imaging phantom 03 medical and health sciences 0302 clinical medicine Optics Position (vector) 030220 oncology & carcinogenesis 0103 physical sciences Radiology Nuclear Medicine and imaging Development (differential geometry) business Proton therapy |
Zdroj: | Physics of Particles and Nuclei Letters. 15:977-980 |
ISSN: | 1531-8567 1547-4771 |
Popis: | In this paper we report a result of theoretical studies of the method for real-time monitoring of the Bragg peak position in a water phantom during scanned proton pencil-beam irradiation. This method is based on the detection of the prompt gamma rays emitted orthogonally to the beam direction produced as a result of inelastic nuclear interactions of primary particles. The principal parameters of the clinical setup prototype and the accuracy of determining the longitudinal coordinates of the Bragg peak position were found on the basis of statistical simulation using the RTS&T Monte Carlo multiparticle transport code. |