Autor: |
Christian M. Langton, Shadi Khoei, Jamie Trapp |
Rok vydání: |
2013 |
Předmět: |
|
Zdroj: |
Journal of Physics: Conference Series. 444:012083 |
ISSN: |
1742-6596 |
DOI: |
10.1088/1742-6596/444/1/012083 |
Popis: |
In this work we used a 3D quantitative CT ultrasound imaging system to characterise polymer gel dosimeters. The system comprised of two identical 5 MHz 128 element phased-array ultrasound transducers co-axially aligned and submerged in water as a coupling agent. Rotational and translational movement of the gel dosimeter sample between the transducers were performed using a robotic arm. Ultrasound signals were generated and received using an Olympus Omniscan unit. Dose sensitivity of attenuation and time of flight ultrasonic parameters were assessed using this system. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|