Characterization of the composite bolometer with a high-Tc superconductor thermometer for an absolute radiometer of synchrotron radiation

Autor: V. Yu. Zerov, Yu. V. Kulikov, Valery F. Pindyurin, I. A. Khrebtov, V. G. Malyarov, D.A. Khokhlov, A. D. Nikolenko, K. V. Ivanov
Rok vydání: 2007
Předmět:
Zdroj: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 575:272-275
ISSN: 0168-9002
DOI: 10.1016/j.nima.2007.01.027
Popis: The article describes the structure and calculation modeling of characteristics of the detector for an absolute synchrotron radiation radiometer in the soft X-ray range of 8–0.4 nm. The detector is a bolometer working on the principle of substitution of electrical power for radiation power, with a high-temperature superconductor film as an element, sensitive to heat. The noise-equivalent power is predicted to be down to 4×10 −11 W/Hz 1/2 at 10 Hz.
Databáze: OpenAIRE