Popis: |
It is widely accepted in capillary zone electrophoresis, that the use of alkaline prerinse procedures can improve the reproducibility of migration times and of corrected peak areas. In this study we present a systematic investigation of alkaline and acid preconditioning procedures for anion analysis using indirect UV detection by designed experiments according to the methodology of Taguchi. Four frequently used electroosmotic flow modifiers (diethylenetriamine, hexamethonium hydroxide, tetradecyltrimethylammonium hydroxide, and hexadimethrine hydroxide) were examined. The optimized procedures were evaluated with regard to the necessary preconditioning time compared to the analysis time. Furthermore, it was demonstrated that the optimized preconditioning technique could be applied to other indirect detection based CE systems. For all examined electrolyte systems, relative standard deviations below 0.5% for migration times and below 5% for corrected peak areas (n=20) were achieved using automated peak integration without further manual reprocessing. |