Methodological Background of the Amersham Amerlex Free Thyroxine RIA

Autor: J. E. M. Midgley
Rok vydání: 1982
Předmět:
Zdroj: Nuklearmedizin. 21:179-183
ISSN: 2567-6407
0029-5566
DOI: 10.1055/s-0037-1620576
Popis: This contribution describes the essential rules underlying a valid estimation of FT4 and compares the way in which various methods deviate from ideal behaviour. The Amerlex(R) FT4 RIA test has been validated by several approaches. A practical demonstration of the ability of various methods to correct for differences in T4-binding protein concentration has been made by independent studies in a UK hospital on euthyroid non-pregnant, euthyroid contraceptive and euthyroid pregnant women. The Amersham kit has been shown to correspond closely to theoretical expectations for FT4 values in these groups.