The Use of an Electronic Particle Counter (Celloscope 101) for Counting Leucocytes

Autor: B. Kvarstein
Rok vydání: 1967
Předmět:
Zdroj: Scandinavian Journal of Clinical and Laboratory Investigation. 19:196-202
ISSN: 1502-7686
0036-5513
DOI: 10.3109/00365516709093502
Popis: Kvarstein, B. The Use of an Electronic Particle Counter (Celloscope 101) for Counting Leucocytes. Scand. J. din. Lab. Invest. 19, 196-202 1967. A method for counting leucocytes with an electronic particle counter using a Cetrimide solution as hemolyzing diluent was examined.The electronic particle counter gives reliable results with a coefficient of variation of 4.4 per cent.The leucocyte counts can be done immediately, or at any time within 24 hours with the same result.Rise in temperature led to a slight fall in white cell counts.The procedure did not require siliconized glass tubes.