Preliminary results of a Compton camera based on a single 3D position-sensitive CZT detector

Autor: Lan Zhang, Yilin Liu, Xu-Ming Ma, Yuanjing Li, Jianqiang Fu, Yulan Li
Rok vydání: 2018
Předmět:
Zdroj: Nuclear Science and Techniques. 29
ISSN: 2210-3147
1001-8042
DOI: 10.1007/s41365-018-0483-0
Popis: A Compton camera prototype has been developed using a pixelated CZT detector with 4-by-4 pixels. Signals of the detector are read out by a VASTAT ASIC that is controlled by a self-developed DAQ board. The DAQ software is developed using LabVIEW, and the offline Compton imaging codes are written in C++. The prototype has been successfully calibrated, and its capabilities for source detection, spectroscopy, and Compton imaging have been demonstrated using a Cs-137 source. The angular resolution of the 662 keV line is 36° FWHM for the simple back-projection method and 9.6° FWHM for the MLEM reconstruction method. The system is ready to be extended to 11-by-11 pixels in the future, and a better imaging quality can be expected due to the better relative position resolution.
Databáze: OpenAIRE