Aqueous humor outflow effects of partial thickness channel created by a femtosecond laser in ex-vivo human eyes

Autor: Gautam Chaudhary, Dongyul Chai, Ron Kurtz, Tibor Juhasz
Rok vydání: 2007
Předmět:
Zdroj: SPIE Proceedings.
ISSN: 0277-786X
DOI: 10.1117/12.701416
Popis: The reduced outflow rate caused by the increased resistance through trabecular meshwork (TM) has been thought to be the main reason for elevated intraocular pressure (IOP). It has been demonstrated that femtosecond laser pulses tuned to 1.7 μm wavelength can create the partial thickness channel in the sclera in ex vivo human eyes [1] and aqueous outflow can be increased by these channels in porcine eyes [2]. It was also shown that the outflow rate is reduced over time in ex vivo human eyes [3]. Therefore, the control experiment without laser treatment at the same condition was conducted and showed that outflow was reduced by 1.5 ± 0.8 μl/min at 15mmHg and 1.8 ± 1.0 μl/min at 25mmHg. However, the outflow rate increased by 0.26 μl/min at 15mmHg and 0.15 μl/min at 25mmHg after the partial thickness channel was created, meaning the amount of increased outflow rate might be more than measured considering the outflow reduction in control experiment. We suggest that the femtosecond laser created partial thickness channel can increase the outflow rate and delay the progression of glaucoma.