Wirbelkörpermorphometrie mit DXA/MXA-Geräten -Ein Gerätevergleich

Autor: T Hoja, J. Mews, Dieter Felsenberg, W. Gowin, E. Diessel
Rok vydání: 1997
Předmět:
Zdroj: RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren. 166:140-145
ISSN: 1438-9010
1438-9029
Popis: Purpose The accuracy and reproducibility of morphometric measurements (Morphometric X-ray Absorptiometry = MXA) of vertebrae were determined. The significant difference of the change in height of vertebral bodies in follow-up studies was computed for MXA methods and digitised spinal radiographs as well. Material and methods The measurements were carried out on two new Dual X-ray Absorptiometry (DXA) devices (device A = Expert, Lunar Corp., device B = QDR 2000 Plus, Hologic Inc.). The data were obtained by using the European Spine Phantom (ESP) and lumbar spine specimens. Results The accuracy of vertebral morphometry performed on radiographs is 2.0%, on the device A 2.3%, and on the device B 4.9%. Measurements taken with the ESP showed a reproducibility of 1.0 to 3.0%, whereas measurements of fractured vertebrae resulted in 5.1 to 6.0%. Conclusion The results of the morphometric measurements demonstrate that a reliable fracture analysis in phantoms and specimens is possible. Further in-vivo studies are necessary.