Popis: |
Ultrasound (US) exposures are widely used with in vitro cell systems e.g. in stem cell and tissue engineering research. However, without the knowledge of factors affecting the level of US exposure, the outcome of the biological result may vary from test to test or even be misinterpreted. Thereby, some of the factors affecting in vitro US exposures were studied.The level of US exposure was characterized in standard commercial cell culturing plates. The temperature distributions were measured inside the wells using infrared camera and fine wire thermocouples, and pressure and intensity distributions using a laser vibrometer and a schlieren system. The measurements were made at operating frequency of around 1 MHz with varying temporal parameters and powers (up to 2 W of acoustic power).Heat accumulation between the wells varied up to 40–50% depending on the location of the well on the plate. This well‐to‐well variation was be linked to the activity of reporter plasmid on osteoblastic cells. Similar temperatu... |