Nuclear and High Energy Physics Photomultiplier medicine.medical_specialty medicine.diagnostic_test Physics::Instrumentation and Detectors business.industry Physics::Medical Physics Detector Particle detector Lyso Semiconductor detector Time of flight Silicon photomultiplier Optics medicine Medical physics business Instrumentation Emission computed tomography |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 703:109-113 |
ISSN: | 0168-9002 |
DOI: | 10.1016/j.nima.2012.11.043 |
Popis: | We have developed γ-ray detectors with good time resolution, to supplement positron emission tomography (PET) with time of flight (TOF) for better localization of the positron annihilation. This will improve the quality of medical images and reduce exposure times. We report results of LYSO scintillating crystals (3×3×15 mm3) irradiated with a 22Na source and coupled to silicon photomultipliers (SiPMs) readout with a waveform digitizer. Improved time resolution, 188 ps FWHM, was obtained with the signal trimmed by a clipping capacitance. |