Autor: |
B. T. Tay, C.B. Tan, A.C. Thai, K. P. Ang, Hari Gunasingham, T.C. Aw, P.Y.T. Teo |
Rok vydání: |
1989 |
Předmět: |
|
Zdroj: |
Analytica Chimica Acta. 221:205-214 |
ISSN: |
0003-2670 |
DOI: |
10.1016/s0003-2670(00)81957-1 |
Popis: |
The development of a practical electrochemical system for in-line blood-glucose monitoring requires the combination of several modifications to achieve reliability. The prototype monitor described includes stopped-flow dialysis, a carbon-fibre pre-electrolysis cell, an immobilized glucose-oxidase reactor and a membrane-coated electrode used with square-wave amperometry. Under microprocessor control, the system is well suited for routine monitoring applications with minimal user interaction. Good precision (RSD of 3–4% ) is achieved over extended use. The linear response range for blood glucose is 1–20 mM. |