Fully automated processing of buffy-coat-derived pooled platelet concentrates

Autor: K. Janetzko, Harald Klüter, Hermann Eichler, Geert van Waeg
Rok vydání: 2004
Předmět:
Zdroj: Transfusion. 44:1052-1058
ISSN: 0041-1132
DOI: 10.1111/j.1537-2995.2004.03296.x
Popis: BACKGROUND The OrbiSac device, which was developed to automate the manufacture of buffy-coat PLT concentrates (BC-PCs), was evaluated. STUDY DESIGN AND METHODS In-vitro characteristics of BC-PC preparations using the OrbiSac device were compared with manually prepared BC-PCs. For standard processing (Std-PC, n = 20), four BC-PCs were pooled using 300 mL of PLT AS (PAS) followed by soft-spin centrifugation and WBC filtration. The OrbiSac preparation (OS-PC, n = 20) was performed by automated pooling of four BC-PCs with 300 mL PAS followed by centrifugation and inline WBC filtration. All PCs were stored at 22 degrees C. Samples were withdrawn on Day 1, 5, and 7 evaluating PTL count, blood gas analysis, glucose, lactate, LDH, beta-thromboglobulin, hypotonic shock response, and CD62p expression. RESULTS A PLT content of 3.1 +/- 0.4 x 10(11) (OS-PCs) versus 2.7 +/- 0.5 x 10(11) (Std-PCs, p < 0.05) was found. A CV of 19 percent (Std-PC) versus 14 percent (OS-PC) suggests more standardization in the OS group. At Day 7, the Std-PCs versus OS-PCs showed a glucose consumption of 1.03 +/- 0.32 micro mol per 10(9) PLT versus 0.75 +/- 0.25 micro mol per 10(9) PLT (p < 0.001), and a lactate production of 1.50 +/- 0.86 micro mol per 10(9) versus 1.11 +/- 0.61 micro mol per 10(9) (p < 0.001). The pH (7.00 +/- 0.19 vs. 7.23 +/- 0.06; p < 0.001), pO(2) (45.3 +/- 18 vs. 31.3 +/- 10.4 mmHg; p < 0.01), and HCO(3) levels (4.91 +/- 1.49 vs. 7.14 +/- 0.95 mmol/L; p < 0.001) suggest a slightly better aerobic metabolism within the OS group. Only small differences in CD62p expression was observed (37.3 +/- 12.9% Std-PC vs. 44.8 +/- 6.6% OS-PC; p < 0.05). CONCLUSION The OrbiSac device allows an improved PLT yield without affecting PLT in-vitro characteristics and may enable an improved consistency in product volume and yield.
Databáze: OpenAIRE