The Effect of Misalignment of Induction Coils on Tissue Heating during Wireless Transcutaneous Energy Transfer
Autor: | E. V. Ryabchenko, Eduard A. Mindubaev, A. A. Danilov |
---|---|
Rok vydání: | 2020 |
Předmět: |
Materials science
business.industry Acoustics Energy transfer 0206 medical engineering Biomedical Engineering Tissue heating Medicine (miscellaneous) 02 engineering and technology 020601 biomedical engineering Power (physics) Medical Laboratory Technology Induction coil Thermal Wireless Coaxial business Displacement (fluid) |
Zdroj: | Biomedical Engineering. 54:203-207 |
ISSN: | 1573-8256 0006-3398 |
Popis: | The effect of induction coil misalignment on the heating of surrounding biological tissues was studied for a wireless inductive powering system with 500 mW output power. It was shown that a lateral misalignment of 2-20 mm can cause an increase in the tissue temperature up to 0.4°C compared with the case of coaxial positioning of the coils. Changes in temperature can be caused by a change in power losses in the elements of the system, as well as by displacement of these elements and the mutual influence of their thermal fields. |