Side lobe artifacts on point spread functions can be traced back to (1) fringe visibility variation across the spectrum, (2) errors in sampling instances, and (3) window functions. We demonstrate signal processing methods for correcting for all three of these issues. These methods require a system calibration step. If the systems slowly age, the recalibration step could be performed in the field with a fixtured target. |