Achieving high reproducibility isotope ratios with the Cameca IMS 1270 in the multicollection mode
Autor: | F. Fernandes, M. Schuhmacher, E. de Chambost |
---|---|
Rok vydání: | 2004 |
Předmět: |
Measurement reproducibility
Reproducibility Isotope Silicon Chemistry Electron multiplier Detector Analytical chemistry General Physics and Astronomy Faraday cup chemistry.chemical_element Surfaces and Interfaces General Chemistry Condensed Matter Physics Isotopes of oxygen Surfaces Coatings and Films symbols.namesake symbols |
Zdroj: | Applied Surface Science. :878-882 |
ISSN: | 0169-4332 |
DOI: | 10.1016/j.apsusc.2004.03.157 |
Popis: | Multicollection system allows improvement of measurement reproducibility for isotope ratios on SIMS instruments provided that the other sources of measurement variability are well controlled. This paper presents analytical protocols developed to control the mass fractionation in the collection optical system and the variation of electron multiplier yield. Results for both silicon and oxygen isotope ratios demonstrate that an experimental error as low 0.2 per mil is achievable whenever both isotopes were measured either with Faraday cup detectors or combining one Faraday cup and one electron multiplier (EM). |