Autor: |
J. P. M. Beijers, H. G. Blosser, S. Brandenburg, A. Hobl, Felix Marti, D. Krischel, M. Schillo, J. Vincent, M. Meyer-Reumers, H.-U. Klein, J. Kim, C. Piel, A. Geisler |
Rok vydání: |
2001 |
Předmět: |
|
Zdroj: |
AIP Conference Proceedings. |
ISSN: |
0094-243X |
Popis: |
A cyclotron for proton therapy has to fulfill many requirements set by the specific operational and safety needs of a medical facility and the medical environment. These are for instance high extraction efficiency, high availability and reliability, simple and robust operation. ACCEL Instruments GmbH has refined the design concept of a medical cyclotron for the PSI PROSCAN project with the objective to use this cyclotron as the standard accelerator in complete proton therapy facilities, which ACCEL intends to market. Starting from the design in [1], we have carried out further detail clarifications, optimizations and adaptations to the needs of PSI [2]. The work was performed in a collaboration between ACCEL, NSCL and KVI in view of the requirements from the PSI PROSCAN project. An overview on the design will be given touching on subjects such as the 3D structural analysis of the coil, detailed magnetic modeling for optimization of the inner region and the spiral, optimization of the RF power, optimization of the cryogenic design based on available cryocoolers instead of a liquefaction plant and Monte Carlo simulations to estimate the heat balance produced by neutrons at 4K components. |
Databáze: |
OpenAIRE |