Assessment of WHO 07/202 reference material and human serum pools for commutability and for the potential to reduce variability among soluble transferrin receptor assays

Autor: Alicia N. Lyle, Jeffrey R. Budd, Victoria M. Kennerley, Bianca N. Smith, Uliana Danilenko, Christine M. Pfeiffer, Hubert W. Vesper
Rok vydání: 2023
Předmět:
Zdroj: Clinical Chemistry and Laboratory Medicine (CCLM).
ISSN: 1437-4331
1434-6621
Popis: Objectives The clinical use of soluble transferrin receptor (sTfR) as an iron status indicator is hindered by a lack of assay standardization and common reference ranges and decision thresholds. In 2009, the WHO and National Institute for Biological Standards and Controls (NIBSC) released a sTfR reference material (RM), 07/202, for assay standardization; however, a comprehensive, formal commutability study was not conducted. Methods This study evaluated the commutability of WHO 07/202 sTfR RM and human serum pools and the impacts of their use as common calibrators. Commutability was assessed for six different measurement procedures (MPs). Serum pools were prepared according to updated CLSI C37-A procedures (C37) or non-C37 procedures. The study design and analyses were based on Parts 2 and 3 of the 2018 IFCC Commutability in Metrological Traceability Working Group’s Recommendations for Commutability Assessment. WHO 07/202 and serum pools were used for instrument/assay and mathematical recalibration, respectively, to determine if their use decreases inter-assay measurement variability for clinical samples. Results The WHO 07/202 RM dilutions were commutable for all 6 MPs assessed and, when used for instrument calibration, decreased inter-assay variability from 208 to 55.7 %. Non-C37 and C37 serum pools were commutable for all 6 MPs assessed and decreased inter-assay variability from 208 to 13.8 % and 4.6 %, respectively, when used for mathematical recalibration. Conclusions All materials evaluated, when used as common calibrators, substantially decreased inter-assay sTfR measurement variability. MP calibration to non-C37 and C37 serum pools may reduce the sTfR IMPBR to a greater extent than WHO 07/202 RM.
Databáze: OpenAIRE