Measurement of neutron tissue dose outside reactor shielding

Autor: I.B. Keirim-Markus, V.V. Markelov, V. T. Korneev, L.N. Uspenskii
Rok vydání: 1963
Předmět:
Zdroj: Soviet Atomic Energy. 15:1132-1139
ISSN: 1573-8205
0038-531X
DOI: 10.1007/bf01115935
Popis: RUS instruments, developed by the authors, are described that enable one to measure the flux and tissue dose rate of intermediate neutrons, which make a significant contribution to neutron tissue dose outside reactor shielding. The neutron dose composition was investigated in experiments at the IRT-1000 reactor, and it was shown that it depends essentially on shielding composition. It was established that the neutron tissue dose computed from readings taken with the RPN-1 instrument were actually too low by a factor amounting to one and one half outside water shielding and to five outside concrete shielding.