medicine.diagnostic_test 010401 analytical chemistry Analytical chemistry Linearity Infrared spectroscopy Molar absorptivity 01 natural sciences 0104 chemical sciences 010309 optics Absorbance chemistry.chemical_compound Wavelength Ultraviolet visible spectroscopy chemistry Spectrophotometry 0103 physical sciences medicine Instrumentation Potassium dichromate Spectroscopy |
Zdroj: | Applied Spectroscopy. 40:1156-1161 |
ISSN: | 1943-3530 0003-7028 |
DOI: | 10.1366/0003702864507611 |
Popis: | Estimations of the uncertainty of eight significant values (λ = 235, 257, 313, 322, 323, 345, 346, and 350 nm) of apparent molar absorptivities in the UV region have been obtained for acidic potassium dichromate solutions at pH 2.21–2.25, recommended as Standard Reference Materials (SRMs). The uncertainties include all the usual sources of possible systematic error in a commercial microprocessor-controlled spectrophotometer. The comparison of these values with those provided by the National Bureau of Standards (U.S.A.) indicates that a precision of ±0.40% has been obtained throughout the absorptivity range. Likewise, the uncertainty in the linearity of experimental absorbances for the eight wavelengths has been estimated, and a table has been drawn up of the absorptivity values and their uncertainties at 10-nm increments over the wavelength range of 200 to 500 nm. |