Anatomic Monopolar Press-fit Radial Head Arthroplasty; High Rate of Loosening at Mid-Term Follow Up

Autor: Inderpaul Samra, Tariq A Kwaees, Wael Mati, Clare Blundell, Suzanne Lane, John WK Harrison, Charalambos P Charalambous
Rok vydání: 2022
Předmět:
Zdroj: Shoulder & Elbow. 15:207-217
ISSN: 1758-5740
1758-5732
DOI: 10.1177/17585732221080768
Popis: Introduction Radial head arthroplasty (RHA) is used for the management of unstable or unreconstructable injuries of the radial head. Our aim was to investigate clinical and radiographic outcomes in patients treated with the Acumed anatomic radial head press-fit system for trauma. Methods Clinical and radiographic assessment of RHAs undertaken for trauma with minimum 2-year follow-up. Results 16 consecutive patients, mean age 53 (21–82) and 66 month ± 27 (26–122) clinical follow-up were included. There were marked radiographic changes with 11/16 showing periprosthetic lucent lines and 13/16 showing subcollar osteolysis. Radiographic changes occurred early post-surgery. Stem loosening was associated with larger cantilever quotients (0.47 vs 0.38, p = 0.004). Overall survivability was 81.2%, with 3 RHAs removed. Clinical outcomes for the retained RHAs were acceptable with mean flexion 134°, extension deficit of 10°, pronation of 82°, and supination of 73°. Mean VAS scores were 8.5 ± 14.4, QuickDASH 13.8 ± 18.9, Mayo Elbow Performance Scores were 91.5 ± 12.5 with no poor scores. Conclusion Mid-term clinical functional outcomes following the Acumed anatomic RHA are acceptable in most cases. However, in view of the extensive periprosthetic lucencies and surgical removal due to loosening, patients should be cautioned when consented for implantation of the prosthesis, especially if a large collar is anticipated.
Databáze: OpenAIRE