Popis: The characterization of a scintillating GEM based gas detector for quality control of clinical radio-therapeutic beams is presented. Photons emitted by the Ar/CF4 gas mixture are detected by means of a CCD camera; in addition, the charge is measured. The detector response has been studied as a function of alpha particle energy and dose rate. The measured signal underestimation, at the Bragg peak depth, is only few percent with respect to an air filled ionization chamber.