Comparison between eight- and sixteen-channel TEM transceive arrays for body imaging at 7 T
Autor: | J. T Vaughan, Greg Metzger, Kamil Ugurbil, Patrick J. Bolan, Jinfeng Tian, Carl Snyder, Steen Moeller, Pierre-Francois Van de Moortele, Can Akgun, Lance DelaBarre |
---|---|
Rok vydání: | 2011 |
Předmět: |
Physics
medicine.diagnostic_test business.industry Power deposition Finite-difference time-domain method Specific absorption rate Magnetic resonance imaging Field strength Nuclear magnetic resonance Optics Male pelvis medicine Radiology Nuclear Medicine and imaging Parallel imaging business Stripline |
Zdroj: | Magnetic Resonance in Medicine. 67:954-964 |
ISSN: | 0740-3194 |
DOI: | 10.1002/mrm.23070 |
Popis: | Eight- and sixteen-channel transceive stripline/TEM body arrays were compared at 7 T (297 MHz) both in simulation and experiment. Despite previous demonstrations of similar arrays for use in body applications, a quantitative comparison of the two configurations has not been undertaken to date. Results were obtained on a male pelvis for assessing transmit, signal to noise ratio, and parallel imaging performance and to evaluate local power deposition versus transmit B(1) (B(1) (+) ). All measurements and simulations were conducted after performing local B(1) (+) phase shimming in the region of the prostate. Despite the additional challenges of decoupling immediately adjacent coils, the sixteen-channel array demonstrated improved or nearly equivalent performance to the eight-channel array based on the evaluation criteria. Experimentally, transmit performance and signal to noise ratio were 22% higher for the sixteen-channel array while significantly increased reduction factors were achievable in the left-right direction for parallel imaging. Finite difference time domain simulations demonstrated similar results with respect to transmit and parallel imaging performance, however, a higher transmit efficiency advantage of 33% was predicted. Simulations at both 3 and 7 T verified the expected parallel imaging improvements with increasing field strength and showed that, for a specific B(1) (+) shimming strategy used, the sixteen-channel array exhibited lower local and global specific absorption rate for a given B(1) (+) . |
Databáze: | OpenAIRE |
Externí odkaz: |