Comparison and improvement of the determinations of actinide low activities using several alpha liquid scintillation spectrometers

Autor: Dacheux, N., Aupiais, J., Courson, O., Aubert, C.
Přispěvatelé: Institut de Physique Nucléaire d'Orsay (IPNO), Université Paris-Sud - Paris 11 (UP11)-Institut National de Physique Nucléaire et de Physique des Particules du CNRS (IN2P3)-Centre National de la Recherche Scientifique (CNRS), Robert, Suzanne
Jazyk: angličtina
Rok vydání: 2000
Předmět:
Zdroj: Analytical Chemistry
Analytical Chemistry, American Chemical Society, 2000, 72, pp.3150-3157
ISSN: 0003-2700
1520-6882
Popis: We applied three procedures using two or liquid scintillation spectrometers (PERALS and TRI-CARB) and two scintillation cocktails (Alphaex and Ultima Gold LLT) for the determination of a-emitter low activities. For each procedure, the limit of detection, the resolution, the separation factor, and the Fischer coefficient were determined in order to perform U-232-U-234-U-238 isotopic measurements. The deconvolution usually performed is clean when the PERALS spectrometer is used. This is not possible for the TRI-CARB spectrometer using the Ultima Gold LLT scintillation cocktail. This problem was solved by combining the advantages of both techniques using the Alphaex scintillation cocktail in the TRI-CARB spectrometer, Under these conditions, the limit of detection was improved, the resolution decreased from 500-800 to 420-590 keV, and the separation factor increased from 0.9 to 1.1-1.2. This third procedure was applied with success for U-232-U-234-U-238 isotopic experiments.