Coincidence Tests of a Compton Telescope Based on Continuous LaBr3 Crystals and SiPMs for Dose Monitoring in Hadron Therapy

Autor: Barrio, J., Cabello, J., Callier, S., Gillam, J., De La Taille, C., Lacasta, C., Rafecas, M., Raux, L., Solaz, C., Stankova, V., Torres-Espallardo, I., Llosa, G.
Přispěvatelé: Instituto de Fisica Corpuscular (IFIC), Consejo Superior de Investigaciones Científicas [Madrid] (CSIC)-Universitat de València (UV), Laboratoire de l'Accélérateur Linéaire (LAL), Centre National de la Recherche Scientifique (CNRS)-Institut National de Physique Nucléaire et de Physique des Particules du CNRS (IN2P3)-Université Paris-Sud - Paris 11 (UP11), Université Paris-Sud - Paris 11 (UP11)-Institut National de Physique Nucléaire et de Physique des Particules du CNRS (IN2P3)-Centre National de la Recherche Scientifique (CNRS)
Jazyk: angličtina
Rok vydání: 2011
Předmět:
Zdroj: 2011 IEEE Nuclear Science Symposium Conference Record
Workshop on Room-Temperature Semiconductor X-ray and Gamma-Ray Detectors, IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011
Workshop on Room-Temperature Semiconductor X-ray and Gamma-Ray Detectors, IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2011, Oct 2011, Valencia, Spain. pp.3539-3542, ⟨10.1109/NSSMIC.2011.6153664⟩
DOI: 10.1109/NSSMIC.2011.6153664⟩
Popis: International audience; In hadron therapy, a proton or carbon ion beam interacts with the tissue and prompt gamma rays are emitted from the excited nuclei that can be employed for dose monitoring. A Compton telescope makes it possible to locate the origin of these photons, allowing an indirect estimation of the delivered dose. The telescope prototype being developed at IFIC-Valencia within the ENVISION project will consist of three detector layers. Each layer will be composed of a continuous LaBr3 crystal coupled to four Silicon Photomultiplier (SiPM) arrays. Coincidence tests have been carried out with a first layer of a 16 mm × 18 mm × 5 mm LaBr3 crystal coupled to a 16 mm × 18 mm SiPM array and a second layer of a 12 mm × 12 mm × 5 mm LYSO crystal coupled to another 16 mm × 18 mm SiPM array. The test system is read out by two SPIROC1 boards working in time coincidence. First data have been taken and the results demonstrate the feasibility of operating the system.