Imaging in Terms Apparent Diffusion Coefficient (ADC) and

Autor: Herek, D, Karabulut, N, Kocyigit, A, Yagci, AB
Jazyk: angličtina
Rok vydání: 2016
Předmět:
Popis: Background: Our aim was to compare the apparent diffusion coefficient (ADC) values of normal abdominal parenchymal organs and signal-to-noise ratio (SNR) measurements in the same patients with breath hold (BH) and free breathing (FB) diffusion weighted imaging (DWI). Material/Methods: Forty-eight patients underwent both BH and FB DWI. Spherical region of interest (ROI) was placed on the right hepatic lobe, spleen, pancreas, and renal cortices. ADC values were calculated for each organ on each sequence using an automated software. Image noise, defined as the standard deviation (SD) of the signal intensities in the most artifact-free area of the image background was measured by placing the largest possible ROI on either the left or the right side of the body outside the object in the recorded field of view. SNR was calculated using the formula: SNR=signal intensity (SI)((organ)/)standard deviation (SD)((noise)). Results: There were no statistically significant differences in ADC values of the abdominal organs between BH and FB DWI sequences (p>0.05). There were statistically significant differences between SNR values of organs on BH and FB DWIs. SNRs were found to be better on FB DWI than BH DWI (p