Urea-based recirculation validation of the symmetrical palindrome catheter
Autor: | Jonczyk, M, Althoff, C, Slowinski, T, Lieker, I, Naik, M, Auer, J, Marcos, RG, Duka, E, Hamm, B, de Bucourt, M |
---|---|
Rok vydání: | 2017 |
Předmět: | |
Zdroj: | Journal Of Renal Care r-IIS La Fe. Repositorio Institucional de Producción Científica del Instituto de Investigación Sanitaria La Fe instname |
ISSN: | 1755-6678 |
Popis: | BACKGROUND: High blood flow and low recirculation rates are central for adequate haemodialysis. A new symmetrical tip has been invented promising efficient haemodialysis even if the ports are reversed. OBJECTIVE: To evaluate access recirculation of the 'palindrome' catheter and to report initial experiences in a clinical setting. MATERIAL AND METHODS: After implantation of the new catheter in 20 patients (male: 14; female: 6; mean age 72 +/- 12.2), access recirculation was evaluated using the urea-based recirculation test. After 30 minutes of haemodialysis, ultrafiltration was stopped and arterial and venous samples were taken. Afterwards, the blood flow rate was reduced to 120 ml/min. Another systemic arterial blood sample was taken 10 seconds after the blood pump was switched off. RESULTS: All 20 interventions were performed successfully without complications. The average recirculation rate was 8.1% with a median of 2.5% ranging from 0 to 85.8%. Recirculation rates under 5% were measured in 13 patients and more than 10% recirculation were found in two patients. The median of days between catheter implantation and recirculation assessment was the day following implantation. CONCLUSION: The new symmetrical catheter presented low recirculation rates in a clinical setting. Since there is just a single tip, fluoroscopic placement in the right atrium is facilitated. |
Databáze: | OpenAIRE |
Externí odkaz: |