Abstrakt: |
ABSTRACTThe present study investigated the feasibility of automating the specimen-pipetting component of sample preparation in the LCx Chlamydia assay (LCx-CT assay; Abbott Laboratories, Chicago, Ill.) by using a commercially available liquid-handling system (Tecan Genesis RSP100; Tecan Inc., Research Triangle Park, N.C.). The Tecan instrument proved to be comparable in both precision and accuracy to a manual multipipettor (Eppendorf model 4850; Eppendorf Scientific, Westbury, N.Y.). The Tecan instrument was extensively checked for evidence of specimen-to-specimen transfer, and no level of contamination sufficient to generate a signal above the background in the LCx-CT assay was detected. Finally, pipetting speed was significantly improved by using the Tecan instrument. A mean time of 2.5 min was required to pipette a complete LCx-CT assay carousel (20 samples and 4 controls) with the Tecan instrument, whereas 8.4 min was required to pipette a comparable number of samples manually (P< 0.001). |