Loomis, T. C., Keith, H. D. |
Zdroj: |
Applied Spectroscopy; July 1975, Vol. 29 Issue: 4 p316-322, 7p |
Abstrakt: |
Direct methods have been devised for calibrating the quantum efficiencies of two x-ray detectors: a Siemens flow-proportional counter in the energy range 2.6 to 11.2 keV, and a General Electric SPG 5 scintillation counter in the range 5.9 to 50.4 keV. Literature values for x-ray attenuation coefficients of detector components are used only to estimate small corrections for x-ray scattering. The largest known errors in the measurements are statistical errors in photon counting and range from 0.03 to 0.23%. In the energy range 5.9 to 11.2 keV, efficiencies of a lithium-drifted silicon detector measured by comparison with the two calibrated detectors as references agree satisfactorily. Linear attenuation coefficients have been measured for argon, methane and P-10 gas (90% Ar; 10% CH4) in the energy range 2 to 12 keV; these can be used to calculate the efficiency of any flow-proportional counter of known geometry when used with counter gas consisting of a known mixture of argon and methane. |