A High-Resolution Portable Gamma-Camera for Estimation of Absorbed Dose in Molecular Radiotherapy

Autor: Bossis, T., Verdier, M.-A., Trigila, C., Pinot, L., Bouvet, F., Blot, A., Ramarijaona, H., Beaumont, T., Broggio, D., Caselles, O., Zerdoud, S., Menard, L.
Zdroj: IEEE Transactions on Radiation and Plasma Medical Sciences; 2024, Vol. 8 Issue: 5 p556-570, 15p
Abstrakt: Molecular radiotherapy is a treatment modality that requires personalized dosimetry for efficient treatment and reduced toxicity. Current clinical imaging systems and miniaturized gamma-cameras lack the necessary features for this task. In this article, we present the design and optimization of a mobile gamma-camera with a $10\times 10$ cm2 field of view tailored for quantitative imaging during $^{131}\text{I}$ therapy of thyroid diseases. The camera uses a monolithic $10\times 10\times 1$ cm3 CeBr3 scintillator coupled to a $16\times 16$ SiPMs array and commercial electronics. It exhibits high imaging performance with an intrinsic spatial resolution (SR) of 1.15-mm FWHM, an energy resolution of 8% FWHM at 356 keV and negligible deadtime up to 150 kcps. Images are reconstructed in real time using a convolutional neural network. The manufacturing method of tungsten collimators and shielding was optimized using laser 3-D printing to achieve an effective density of 97% that of bulk tungsten. Their geometry was adjusted with Monte-Carlo simulations in order to reduce septal penetration and scattering and optimize the signal-to-noise ratio at short times after treatment administration. Two high-energy parallel-hole collimators with high sensitivity or very high SR were designed for treatment planning and post-treatment control. The fully operational gamma-camera will soon be clinically assessed.
Databáze: Supplemental Index