Autor: |
Biloivan, Olga A., Dzyadevych, Sergiy V., Boubriak, Olga A., Soldatkin, Alexey P., El'skaya, Anna V. |
Zdroj: |
Electroanalysis; November 2004, Vol. 16 Issue: 22 p1883-1889, 7p |
Abstrakt: |
A new potentiometric biosensor allowing quantitative determination of the proteinases through their esterase activity has been developed. The biosensor, specific for ester artificial substrates of serine proteinases has been fabricated using a pH‐sensitive field effect transistor (pH‐FET) and an immobilized complex of trypsin and α2‐macroglobulin. It has been shown that created biosensor is able to determine the activity of the soluble trypsin in the range of 0.1–30 U/mL (substrate BAEE). The relative standard deviation for the trypsin determination was approximately 3%. The operation stability of the biosensor was no less than 12 h (40 measurements). The response of the biosensor stored at +4 °C was stable for 30 days. |
Databáze: |
Supplemental Index |