Autor: |
Alberts, Melissa S., Niles, Scott D., Wilson, Dennis, Ploessl, James, Alberts, Melissa S., Niles, Scott D., Wilson, Dennis, Ploessl, James |
Zdroj: |
Journal of Extra-Corporeal Technology; June 2009, Vol. 41 Issue: 2 p105-109, 5p |
Abstrakt: |
Triboelectric charging is commonly detected during cardiopulmonary bypass in circuits using roller pumps and PVC tubing. Dissipation of this charging is needed to prevent a spontaneous discharge from occurring. We evaluated the ability of the Quadrox-D (Maquet) to effectively remove the electrostatic charge accumulation, with and without a heater/cooler connection (H/C). A Quadrox-D oxygenator was evaluated using a Stockert SIII pump head, Cincinnati subzero H/E, and a custom adult extracorporeal membrane oxygenation tubing pack with 1/2′ PVC raceway. The circuit was primed and evaluated for triboelectric accumulation with and without H/C use. The results showed a linear relationship between increasing pump flow and static charge build-up when an H/C was not applied. The calculated r2value was .95. Incorporation of the H/C effectively eliminated charge accumulation. Increasing pump speed increases the amount of static charge created without the use of an H/C. Incorporation of an H/C effectively eliminates charge build-up in the Quadrox-D and is recommended while priming the circuit. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|