Autor: |
Henson, J. H. L, Jeans, W. D., Newham, F. J., Paice, C., Blakeborough, A. |
Zdroj: |
The British Journal of Radiology; November 1988, Vol. 61 Issue: 731 p1026-1031, 6p |
Abstrakt: |
Gruntzig balloon catheters from three manufacturers have been tested in vitroto establish a method of assessing the diameters and bursting pressures of inflated balloons. The method of testing and of measuring the inflated balloons is described. It was found that there were considerable variations in the behaviour of different balloons, and the relevance of this to arterial rupture is discussed. There were significant associations between the manufacturer and premature rupture of the balloon (p< 0.01), and between co-axial construction of the catheter and overdistension of the balloon by more than 20% (p< 0.01). The balloon length and the stated maximum diameter were not associated with either premature failure or overdistension. |