Autor: |
Eddleston, B., Moores, B. M., Walker, A. |
Zdroj: |
The British Journal of Radiology; April 1977, Vol. 50 Issue: 592 p266-270, 5p |
Abstrakt: |
Eye dose measurements have been performed when using medium speed conventional and rare-earth screen-film combinations during tomography of the para-nasal sinuses. The measurements show that using conventional intensifying screens with the A.P. view a total eye dose of about 20 rad may be given during an examination. This eye dose can be reduced by 98% using the P.A. position. If rare-earth screen/film combinations are employed the eye dose measured in the A.P. view was reduced by 75% of that obtained with conventional screens without detectable loss of image quality. A total eye dose reduction of about 99·5% was measured in the P.A. view with the rare-earth systems. |