Autor: |
Koenig, S C, VanHimbergen, D J, Jaber, S F, Ewert, D L, Cerrito, P, Spence, P A |
Zdroj: |
European Journal of Cardio-Thoracic Surgery; September 1999, Vol. 16 Issue: Supplement 1 pS83-S87, 5p |
Abstrakt: |
Flow probes have been introduced as a non-invasive means of anastomotic quality assessment in off-pump coronary artery bypass graft (CABG). Flow waveform morphology cannot reliably be assessed visually unless severe anastomotic stenosis is present ( > 90%). We applied spectral analysis techniques to determine whether the frequency content of graft flow can improve the surgeon's ability to detect anastomotic errors. |