Autor: |
Liu, Lin, Hussain, Syeda K., Miller, Robert S., Oyarzabal, Omar A. |
Zdroj: |
Journal of Food Protection; November 2009, Vol. 72 Issue: 11 p2428-2432, 5p |
Abstrakt: |
The goals of this study were to evaluate the efficacy of the mini VIDAS automated immunoassay chemistry system to detect Campylobacterspp. from retail broiler meat enriched in Bolton broth supplemented with lysed blood (B+B) or without blood (B–B), and to detect positive samples at 24 versus 48 h after enrichment. Retail broiler meat was enriched and tested for Campylobacterspp. with the mini VIDAS and with an agar plate. Isolates were speciated with a multiplex PCR and typed with pulsed-field gel electrophoresis (PFGE) to evaluate relatedness of isolates collected from subsamples enriched in B+B or B–B. The number of Campylobacter-positive samples by mini VIDAS was similar (P> 0.05) to the results found with traditional plating media for naturally contaminated broiler meat, regardless of whether the comparison was made between B+B and B–B, or among different meat products (breast, tenders, and thighs). More positive samples were found at 48 h of enrichment than at 24 h of enrichment (P< 0.05). A Campylobacter jejuni:Campylobacter coliratio of 4:1 was found in this study. Most of the isolates from both subsamples (B+B and B–B) were similar or identical by PFGE analysis, except for a few samples in which the PFGE profiles of the isolates from the subsamples were different. Mini VIDAS allowed for the detection of Campylobacterspp. within 48 h after enrichment. However, the sensitivity is similar to plate media, and retail broiler samples need to be enriched for 48 h to avoid false negatives. |