Autor: |
Yanase, Nobuyuki, Naganawa, Hirochika, Nagano, Tetsushi, Noro, Junji |
Zdroj: |
Analytical Sciences; March 2011, Vol. 27 Issue: 3 p325-328, 4p |
Abstrakt: |
A single current “emulsion flow” liquid-liquid extraction apparatus has a head with a number of holes from which micrometer-sized droplets of an aqueous phase spout into an organic phase to mix the two liquid phases. For practical use, however, a fatal problem can occur when particulate components in the aqueous phase plug the holes. In the present study, we have succeeded in solving the problem by applying a counter current-type emulsion flow extractor where micrometer-sized droplets of the organic phase are generated. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|