Autor: |
Sha, Jian, Endsley, Janice J., Kirtley, Michelle L., Foltz, Sheri M., Huante, Matthew B., Erova, Tatiana E., Kozlova, Elena V., Popov, Vsevolod L., Yeager, Linsey A., Zudina, Irina V., Motin, Vladimir L., Peterson, Johnny W., DeBord, Kristin L., Chopra, Ashok K. |
Zdroj: |
Journal of Clinical Microbiology; May 2011, Vol. 49 Issue: 5 p1708-1715, 8p |
Abstrakt: |
ABSTRACTWe evaluated two commercial F1 antigen capture-based immunochromatographic dipsticks, Yersinia Pestis (F1) Smart II and Plague BioThreat Alert test strips, in detecting plague bacilli by using whole-blood samples from mice experimentally infected with Yersinia pestisCO92. To assess the specificities of these dipsticks, an in-frame F1-deficient mutant of CO92 (?caf) was generated by homologous recombination and used as a negative control. Based on genetic, antigenic/immunologic, and electron microscopic analyses, the ?cafmutant was devoid of a capsule. The growth rate of the ?cafmutant generally was similar to that of the wild-type (WT) bacterium at both 26 and 37°C, although the mutant's growth dropped slightly during the late phase at 37°C. The ?cafmutant was as virulent as WT CO92 in the pneumonic plague mouse model; however, it was attenuated in developing bubonic plague. Both dipsticks had similar sensitivities, requiring a minimum of 0.5 µg/ml of purified F1 antigen or 1 × 105to 5 × 105CFU/ml of WT CO92 for positive results, while the blood samples were negative for up to 1 × 108CFU/ml of the ?cafmutant. Our studies demonstrated the diagnostic potential of two plague dipsticks in detecting capsular-positive strains of Y. pestisin bubonic and pneumonic plague. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|