Autor: |
Smith, P B, Rhoden, D L, Tomfohrde, K M |
Zdroj: |
Journal of Clinical Microbiology; April 1975, Vol. 1 Issue: 4 p359-362, 4p |
Abstrakt: |
The PathoTec Rapid I-D System for identifying Enterobacteriaceae was evaluated with 471 cultures. In 4,910 individual test comparisons, 95.5% of the results agreed, with results of only two test strips, those for esculin hydrolysis and urease production, agreeing with conventional tests in less than 94% of the trials. The PathoTec system exhibited 94.3% accuracy in identifying these cultures in a double-blind study with conventional media and procedures as the alternate system. Two newly developed test strips, for 0-nitrophenyl-beta-D-galactopyranoside and ornithine decarboxylase, were found to be highly reliable. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|