Abstrakt: |
The advantage of high-speed time-of-flight-mass spectrometry (TOF-MS) detection for ultrafast qualitative supercritical fluid chromatography/mass spectrometry (SFC/MS) applications allows the superior resolving power of SFC to be exploited in high-throughput analysis. A chromatographic comparison of quadrupole MS and TOF-MS shows high-speed TOF total ion current data point sampling to be more indicative of fast SFC separations and corresponding short (1−2 s) baseline peak widths. Results shown for analysis of a six-compound mixture with two peaks eluting at 0.86 and 0.89 min exhibit >50% resolution by high-speed TOF data sampling, whereas the same peaks appear to coelute using quadrupole MS data sampling. Additionally, a marked improvement in the peak baseline widths is afforded by fast TOF data acquisition of 0.1 s/spectrum, resulting in a reduction in the baseline width, 1.6 s, of sulfanilamide in a four-compound mixture that is more than 2-fold greater than that achieved at the slower data acquisition of 0.5 s/spectrum. The resulting increase in resolution and improved peak shapes allow automatic integration routines to perform more effectively. For most classes of compounds amenable to high performance liquid chromatography, including druglike species, steroids, and polymers, the union of SFC with TOF-MS provides the maximum density of chemical information per unit time available with any high-speed chromatographic/mass spectrometric method. |