HPLC electrochemical detection and quantification of monoamines and their metabolites in rat brain tissue samples

Autor: Viljoen, F. P., Preez, J. L. du, Wessels, J. C., Aucamp, M. E.
Zdroj: Die Pharmazie; October 2018, Vol. 73 Issue: 10 p563-569, 7p
Abstrakt: The monitoring of monoamines and their metabolites in CNS samples can be very valuable in pharmaceutical and biomedical research. A specific high performance liquid chromatography, coupled to a coulometric electrochemical detection method, for the assay of monoamines (dopamine, norepinephrine, epinephrine and serotonin) and their metabolites in rat brain tissue samples was developed. The chromatographic separation was achieved on a C8 reversed phase column with a mobile phase consisting of 0.1 M sodium formate buffer, 5 mM sodium 1-heptanesulfonate, 0.17 mM ethylenediaminetetraacetic acid disodium salt and 5% v/v acetonitrile (pH ±4.0). The detection was achieved through electrochemical detection, with a coulometric cell potential setting of +650 mV. The flow-rate was at 1 ml/min and the total run time was 50 min. The method was validated according to validation guidelines. The method was found to be linear (R2> 0.99) over the analytical range (5 to 200 ng/ml) for all monoamines and their metabolites. All the other validation parameters were acceptable and within range. The method was applied to three rat brain areas (pre-frontal cortex, hippocampus and striatum), where the monoamines (except for epinephrine) and their metabolites were easily detected.
Databáze: Supplemental Index